CommissioningIntermediateMeasurement

Instrument Uncertainty Calculator

Combine accuracy, resolution, and an additional independent uncertainty component by root-sum-square.

Why it matters: A measured difference should not be over-interpreted when it is comparable to the combined measurement uncertainty.

User inputs

Enter field measurements

5 inputs
selected unitsMagnitude of the reported reading.
% of readingEnter a standard uncertainty component expressed as percent of reading. Convert a published accuracy limit before using it here.
selected unitsSmallest display increment; half a digit is used as a rectangular screening component.
selected unitsAdditional standard uncertainty from method, repeatability, or environment.
%Project- or test-procedure-specific maximum. Replace the editable example value; no universal acceptance limit is assumed.

Calculated result

Engineering evaluation

Within entered screening criteria
Combined relative standard uncertainty (Ur)1.118 %
Entered percent-of-reading standard uncertainty component (Ua)
1 selected units
Resolution component (Ures)
0.029 selected units
Combined standard uncertainty (U)
1.118 selected units

Engineering interpretation

The combined standard uncertainty is 1.118 selected units, or 1.118% of the reading, compared with the entered 2% relative-uncertainty limit. This screening combination assumes independent standard components and does not automatically provide expanded uncertainty or a coverage interval.

Important limitations

  • The percent-of-reading input is treated as a standard uncertainty component. Convert a published accuracy limit to an appropriate standard uncertainty before formal analysis, and verify the entered project criterion.

Engineering knowledge center

Understand the calculation—not just the answer

2026-08-07 · Free Library
Governing relationships
Ua = |X|A/100
Ures = R/√12
U = √(Ua²+Ures²+Uo²)
Ur = U/|X|×100
within entered criterion when Ur ≤ L
Current substitution

Ua = |100|×1/100 = 1; Ures = 0.1/√12 = 0.02887; U = 1.118; Ur = 1.118%.

Calculation sequence

Step-by-step method

  1. Convert the entered percent-of-reading standard uncertainty to an absolute component.
  2. Represent digital resolution as a rectangular standard uncertainty.
  3. Combine independent standard uncertainty components by root-sum-square and divide by reading magnitude.

Equation legend

Inputs and calculated quantities

X
Measured reading — selected units
A
Percent-of-reading standard uncertainty component — % of reading
R
Display resolution — selected units
Uo
Other independent standard uncertainty — selected units
L
Entered maximum relative uncertainty — %
Ur
Combined relative standard uncertainty — %
Ua
Entered percent-of-reading standard uncertainty component — selected units
Ures
Resolution component — selected units
U
Combined standard uncertainty — selected units

Engineering calculation disclaimer

These engineering tools are provided for educational, preliminary evaluation, field verification, and commissioning support. Results depend on the accuracy of user-entered information and the assumptions stated for each calculation. Every colored status and comparison is a screening indicator only; project criteria, contract requirements, applicable codes, manufacturer instructions, measurement uncertainty, and authorized engineering judgment govern. The tools do not replace project-specific engineering analysis or the judgment of a licensed professional engineer.